Question 1 · Hard
A patient with a programmable ventriculoperitoneal shunt valve is scanned at 1.5 T under the valve's MR conditional labelling. What must be arranged after the scan?
- AA CT is done to confirm that the valve has not moved
- BThe shunt is tapped to test for infection
- CThe valve setting is checked and reset if needed
- DNo action, as an MR conditional scan needs no follow-up
Show answerHide answer
Answer: C · The valve setting is checked and reset if needed
Many programmable shunt valves are set by an external magnet, so the scanner's field can change the pressure setting. Their labelling allows scanning, but the setting must be checked, usually with a radiograph or a reading device, and reset by the neurosurgical team after MRI. An unnoticed change can cause over- or under-drainage of CSF.
- A: The concern is a changed setting, not movement of the valve.
- B: Tapping the shunt is an invasive test for infection and is not part of MRI care.
- D: MR conditional means safe under conditions, which here include checking the setting afterwards.
Remember this
Check and reset programmable shunt valves after MRI.